Blood Glucose Device Linearity

XMLWordPrintable

    • Type: Hosp Inpt eCQMs - Hospital Inpatient eCQMs
    • Resolution: Resolved
    • Priority: Moderate
    • Component/s: None
    • None
    • Beth Sahlmann
    • 814-449-2547
    • VHA
    • Hide
      Thank you for your inquiry regarding CMS871v5/HH-Hyper, Hospital Harm - Severe Hyperglycemia, and CMS816v5/HH-Hypo, Hospital Harm - Severe Hypoglycemia.

       

      The measure specifications evaluate glucose results using the numeric result and the applicable threshold defined in the measure logic. For example, HH-Hypo identifies a glucose result of less than 40 mg/dL when evaluating a potential severe hypoglycemic event. HH-Hyper similarly evaluates glucose results against the thresholds specified in the measure logic when evaluating a hospitalization against measure population and measure observations criteria.

       

      We recognize that some point-of-care glucometers may report results outside of their validated reportable range using a comparator, such as “<30 mg/dL,” “<40 mg/dL,” or “<50 mg/dL,” rather than reporting a discrete numeric glucose value. The measure specifications do not include special logic for handling device-specific reporting conventions or point-of-care device linearity limitations. The result should be represented in accordance with the applicable health information technology and data exchange standards, including the reported comparator and value, as supported.

       

      For purposes of evaluating the measure logic, a result reported with a comparator can only establish whether the applicable measure threshold is satisfied when the reported result definitively establishes that relationship. For example, a result reported as “<30 mg/dL” establishes that the glucose result is less than 40 mg/dL. In contrast, a result reported as “<50 mg/dL” does not establish that the glucose result is less than 40 mg/dL, because the actual glucose concentration could be between 40 and 50 mg/dL.

       

      Therefore, implementers should not create an assumed numeric result solely for measure calculation when the device reports only a bounded result. The reported result and comparator should be retained, and the measure criterion should be considered met only when the reported information is sufficient to determine that the applicable threshold specified by the measure has been met.

      [For follow-up questions submitted more than 30 days after the "Solution Posted On" date, please submit a new Jira ticket.] 
      Show
      Thank you for your inquiry regarding CMS871v5/HH-Hyper, Hospital Harm - Severe Hyperglycemia, and CMS816v5/HH-Hypo, Hospital Harm - Severe Hypoglycemia.   The measure specifications evaluate glucose results using the numeric result and the applicable threshold defined in the measure logic. For example, HH-Hypo identifies a glucose result of less than 40 mg/dL when evaluating a potential severe hypoglycemic event. HH-Hyper similarly evaluates glucose results against the thresholds specified in the measure logic when evaluating a hospitalization against measure population and measure observations criteria.   We recognize that some point-of-care glucometers may report results outside of their validated reportable range using a comparator, such as “<30 mg/dL,” “<40 mg/dL,” or “<50 mg/dL,” rather than reporting a discrete numeric glucose value. The measure specifications do not include special logic for handling device-specific reporting conventions or point-of-care device linearity limitations. The result should be represented in accordance with the applicable health information technology and data exchange standards, including the reported comparator and value, as supported.   For purposes of evaluating the measure logic, a result reported with a comparator can only establish whether the applicable measure threshold is satisfied when the reported result definitively establishes that relationship. For example, a result reported as “<30 mg/dL” establishes that the glucose result is less than 40 mg/dL. In contrast, a result reported as “<50 mg/dL” does not establish that the glucose result is less than 40 mg/dL, because the actual glucose concentration could be between 40 and 50 mg/dL.   Therefore, implementers should not create an assumed numeric result solely for measure calculation when the device reports only a bounded result. The reported result and comparator should be retained, and the measure criterion should be considered met only when the reported information is sufficient to determine that the applicable threshold specified by the measure has been met. [For follow-up questions submitted more than 30 days after the "Solution Posted On" date, please submit a new Jira ticket.] 
    • Not measure related
    • CMS0816v6, CMS0871v6
    • CMS0816v5, CMS0871v5
    • CMS0816v4, CMS0871v4
    • CMS0816v3, CMS0871v3
    • Are there guidelines established for dealing with glucometer device linearity results reported as less than a particular value.

      We are seeing glucose values generated by glucometers that are reported as less than a particular value (e.g., <30,<40, <45, <50, etc.) due to device linearity calculations. Device linearity is a quality-assessment process used to verify that a glucometer provides accurate, proportional glucose readings across its entire reportable range. This calculation is essential for ensuring clinicians can rely on the meter in both hypoglycemia and hyperglycemia ranges. We have created some interim rules for addressing these within our scoring logic but were wondering if the owners of this measure have developed guidelines for how this should be addressed with the measure calculations?